Recommended Conditions
Elite 15x150 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Elite 15x150 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at optimal ...
Ultimate 25x150 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Ultimate 25x150 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...
Rapid 8x75 column – Recommended conditions
Temperature & flow rate We have tested flow rates from 100 to 1000 nl/min and all produced great results. Maximum pressure The Aurora Series column and nanoZero® fitting can both handle pressures >1700 bar. Expected back pressure The below graph ...
Rapid 8x150 column – Recommended conditions
Temperature & flow rate We tested flow rates ranging from 0.4 to 3 μL/min on the Rapid 8×150 column, all of which produced excellent results. At these flow rates and room temperature, the back pressure from the column should remain within normal ...
Elite 15x75 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Elite 15x75 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at optimal ...
Rapid 5x75 column – Recommended conditions
Temperature & flow rate We tested flow rates ranging from 100 to 2000 nl/min on the Rapid 5×75 column, all of which produced excellent results. At these flow rates and room temperature, the back pressure from the column should remain within normal ...
Ultimate 25x75 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Ultimate 25x75 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...
Rapid 5x150 column – Recommended conditions
Temperature & flow rate We tested flow rates ranging from 0.4 to 4 μL/min on the Rapid 5×150 column, all of which produced excellent results. At these flow rates and room temperature, the back pressure from the column should remain within normal ...
Frontier 60x75 column – Recommended conditions
Temperature & flow rate A column oven must be installed at the mass spectrometer source for the 60 cm Aurora Frontier column. For typical operation we recommend a flow rate between 100 and 300 nl/min with the column oven set to 60°C. The maximum ...