Recommended conditions for 25cm Aurora Ultimate column

Recommended conditions for 25cm Aurora Ultimate column

Temperature & flow rate
We recommend the use of a column oven installed at the mass spectrometer source for the 25cm Aurora Ultimate column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at optimal flow rates, generating inconsistent sample separation.

For typical operation we recommend a flow rate around 400 nl/min with the column oven set to 40-60°C. This will ensure UHPLC back pressures of around 600-700 bar. The maximum operating temperature is 60°C. Flow rates can be increased or decreased to suit particular applications.

Maximum pressure
Aurora Series columns can withstand pressures >1700 bar (Gen3) and >1200bar (Gen2).

Example gradient
Refer to this Help Centre article for an example gradient.

Expected back pressure
Refer to this Help Centre article for the expected backpressure of a 25cm Aurora Ultimate column, for a given flow rate at both room temperature and 40°C.


For additional details on the operation of Aurora Series columns, please refer to this Help Centre section.

    • Related Articles

    • Recommended conditions for 60cm Aurora Frontier column

      Temperature & flow rate A column oven must be installed at the mass spectrometer source for the 60cm Aurora Frontier column. For typical operation we recommend a flow rate between 200 and 300 nl/min with the column oven set to 60°C. The maximum ...
    • Recommended conditions for 15cm Aurora Elite column

      Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the 15cm Aurora Elite column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...
    • Recommended conditions for 5cm Aurora Rapid75 column

      Temperature & flow rate We have tested flow rates from 50 to 2000 nl/min and all produced great results. At flow rates of between 50 and 2000 nl/min at room temperature, the back pressure from the column should be within normal operating range of ...
    • Recommended conditions for 5cm Aurora Rapid150 column

      Temperature & flow rate We have tested flow rates from 2 to 4 μl/min and all produced great results. At flow rates of between 2 and 4 μl/min at room temperature, the back pressure from the column should be within normal operating range of most ...
    • Expected back pressure, 25cm Aurora Ultimate column

      The below graph depicts the expected backpressure of a 25cm Aurora Ultimate column, for a given flow rate at both room temperature and 40°C. The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid. The Aurora Series column and ...