Ultimate 25x75 column – Recommended conditions
Temperature & flow rate
We recommend the use of a column oven installed at the mass spectrometer source for the Ultimate 25x75 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at optimal flow rates, generating inconsistent sample separation.
For typical operation we recommend a flow rate between 100 to 400 nl/min with the column oven set between 40-60°C. The maximum operating temperature is 60°C. Flow rates can be increased or decreased to suit particular applications.
Maximum pressure
Aurora Series columns and nanoZero® fitting can both handle pressures can withstand pressures >1700 bar.
Expected back pressure
The below graph depicts the expected backpressure of a Ultimate 25x75 column, for a given flow rate at both room temperature and 40°C. The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid.
See example gradients for the full range of Aurora Series columns here.
Related Articles
Ultimate 25x150 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Ultimate 25x150 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...
Column volumes and equilibration
For initial installation Aurora Series columns should be equilibrated via direct flow with 5 column volumes of 80% buffer B (100% Acetonitrile, 0.1% Formic Acid) followed by 5 column volumes of 1% buffer B. WARNING: Do not run UHPLC system column ...
Rapid 8x75 column – Recommended conditions
Temperature & flow rate We have tested flow rates from 100 to 1000 nl/min and all produced great results. Maximum pressure The Aurora Series column and nanoZero® fitting can both handle pressures >1700 bar. Expected back pressure The below graph ...
Elite 15x150 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Elite 15x150 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at optimal ...
Elite 15x75 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Elite 15x75 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at optimal ...