Rapid 8x150 column – Recommended conditions

Rapid 8x150 column – Recommended conditions

Temperature & flow rate
We tested flow rates ranging from 0.4 to 3 μL/min on the Rapid 8×150 column, all of which produced excellent results. At these flow rates and room temperature, the back pressure from the column should remain within normal operating range of most UHPLC systems. Flow rates above this range typically require column heating to keep back pressure within manageable levels.

Maximum pressure
Aurora Series columns and nanoZero® fitting can both handle pressures can withstand pressures >1700 bar.

Expected back pressure
The below graph depicts the expected backpressure of a Rapid 8x150 column, for a given flow rate at both room temperature and 50°C. The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid.



Example gradient
See example gradients for the full range of Aurora Series columns here.