Recommended conditions for 5cm Aurora Rapid75 column

Recommended conditions for 5cm Aurora Rapid75 column

Temperature & flow rate
We have tested flow rates from 50 to 2000 nl/min and all produced great results. At flow rates of between 50 and 2000 nl/min at room temperature, the back pressure from the column should be within normal operating range of most UHPLCs. Higher flow rates generally require heating to keep the back pressure within manageable levels.

Maximum pressure
Aurora Series columns can withstand pressures >1700 bar (Gen3) and >1200bar (Gen2).

Example gradient
Refer to this Help Centre article for an example gradient.

Expected back pressure
Refer to this Help Centre article for the expected backpressure of a 5cm Aurora Rapid75 column, for a given flow rate at both room temperature and 40°C.


For additional details on the operation of Aurora Series columns, please refer to this Help Centre section.
    • Related Articles

    • Recommended conditions for 60cm Aurora Frontier column

      Temperature & flow rate A column oven must be installed at the mass spectrometer source for the 60cm Aurora Frontier column. For typical operation we recommend a flow rate between 200 and 300 nl/min with the column oven set to 60°C. The maximum ...
    • Recommended conditions for 15cm Aurora Elite column

      Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the 15cm Aurora Elite column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...
    • Recommended conditions for 25cm Aurora Ultimate column

      Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the 25cm Aurora Ultimate column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...
    • Recommended conditions for 5cm Aurora Rapid150 column

      Temperature & flow rate We have tested flow rates from 2 to 4 μl/min and all produced great results. At flow rates of between 2 and 4 μl/min at room temperature, the back pressure from the column should be within normal operating range of most ...
    • Expected back pressure, 15cm Aurora Elite column

      The below graph depicts the expected backpressure of a 15cm Aurora Elite column, for a given flow rate at both room temperature and 40°C. The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid. The Aurora Series column and ...