Operation of Aurora Series columns - removal from a UHPLC

Operation of Aurora Series columns - removal from a UHPLC

If possible, do not remove the column from a UHPLC system before the column performance falls and the columns requires replacement, however, removal for instrument servicing and maintenance is sometimes required.

Removal of the nanoViper™ Fingertight Fitting from the nanoZero® under high back pressure conditions can damage the stationary phase bed and lead to blockages and poor chromatographic performance.
The removal procedure can lead to fouling of the emitter tip and poor column performance. The IonOpticks replacement guarantee is not valid if a column has been removed from a UHPLC once in operation.
For removal of the column for extended periods of time we recommend placing 30 μl of methanol in the nanoZero® and screwing in the plug supplied during transport.

Removal of Aurora and Aurora CSI columns:

  1. Run 80% B for 5min at operating flow rates before reducing flow to 0.002 μl/min for 10min or until the back pressure has stabilised below 10 bar
  2. Set MS system into standby mode
  3. The nanoViper™ Fingertight Fitting can then be removed from the nanoZero®.
  4. Screw the shipping plug into the nanoZero®. For removal of the column for extended periods of time we recommend placing 30μl of methanol in the nanoZero® before installing the shipping plug

Removal of Aurora TS columns from Thermo Scientific's Nanospray Flex source:

  1. Run 80% B for 5min at operating flow rates
  2. Reduce flow to 0.002 μl/min for 10 min or until the back pressure has stabilised below 10 bar
  3. Set MS system into standby mode
  4. Slide Nanospray Flex stage out of the safety hood before sliding the column cassette backwards from the TS Nanospray Flex Adapter along the column holder
  5. The nanoViper™ Fingertight Fitting can now be removed from the nanoZero®
  6. Carefully place the protective cap back onto the cassette and press firmly to ensure that the cap is held securely
  7. Screw the shipping plug into the nanoZero®. For removal of the column for extended periods of time we recommend placing 30μl of methanol in the nanoZero® before installing the shipping plug

Removal of Aurora TS columns from Thermo Scientific's EasySpray source:

  1. Run 80% B for 5min at operating flow rates
  2. Reduce flow to 0.002 μl/min for 10 min or until the back pressure has stabilised below 10 bar
  3. Set MS system into standby mode
  4. Slide column cassette backwards from EasySpray source along the column holder
  5. The nanoViper™ Fingertight Fitting can now be removed from the nanoZero®
  6. Carefully place the protective cap back onto the cassette and press firmly to ensure that the cap is held securely
  7. Screw the shipping plug into the nanoZero®. For removal of the column for extended periods of time we recommend placing 30μl of methanol in the nanoZero® before installing the shipping plug

    • Related Articles

    • Operation of Aurora Series columns - standby and idle conditions

      To optimise column lifetime and performance, we recommend that the following conditions are met while your instrument is standby and idle mode: The instrument continues to run at the desired operating pressure The instrument ideally continues to run ...
    • Operation of Aurora Series columns - initial operation

      Once the column is connected to your UHPLC system and is placed inside the source heater or housing, begin operation using 70% buffer B at a flow rate equivalent to the desired gradient flow rate for around 10 minutes or until the pressure is stable ...
    • Operation of Aurora Series columns - sample loading & buffer compositions

      We recommend the following buffer compositions: Buffer A: 99.9% MilliQ Water, 0.1% formic acid Buffer B: 99.9% Acetonitrile, 0.1% formic acid Please consult with your UHPLC manufacturer to confirm that your UHPLC is compatible with these ...
    • Operation of Aurora Series columns - sample preparation

      Prior to loading samples onto your Aurora Series column, the sample should be de-salted and should not contain any contaminants (salts, detergents, solid particles, etc). Loading contaminated samples onto the column may disrupt solvent flow or foul ...
    • Operation of Aurora Series columns - column equilibration

      Before each run the column should be reequilibrated using a minimum of 4 column volumes of 100% buffer A. Column equilibration should be performed below 1000 bar. Column volumes: Aurora Frontier, 60cm X 75μm: 2.65μl Aurora Ultimate, 25cm X 75μm: ...