How to connect the nanoZero® to the nanoViper™ fitting
How to connect the nanoZero® to the nanoViper™ fitting
This video demonstrates how to connect the nanoZero to the Thermo Scientific nanoViper™ fitting:
Hold the nanoZero with the spanner.
Tighten the nanoViper™ fitting finger tight until you reach a firm stop.
A firm stop is the point at which the nanoViper™ stops turning and significant additional force would be required to turn the fitting further.
Due to the unique design of the nanoZero, there is no bore between the fittings. The end of the nanoViper™ interfaces directly with a polymer inside the nanoZero, creating a true-zero dead volume union.
Pro Tips:
Operating the column with an under-tight fitting can result in a loss of performance that cannot be recovered with additional tightening.
Additionally, we recommend only tighten the fitting finger tight, and not to use a torque tool.
This article demonstrates how to install an Aurora CSI column to your UHPLC system with the Bruker CaptiveSpray source using a Sonation Column Oven. It includes a video tutorial and step-by-step guidance from our User Guide on: Installing the Aurora ...
This article demonstrates how to install an Aurora CSI column to your UHPLC system with Bruker CaptiveSpray source using a Bruker Column Toaster. It includes a video tutorial and step-by-step guidance from our User Guide on: Installing the Aurora CSI ...
<br> This article demonstrates how to install an Aurora Rapid 5 cm column on your UHPLC system with the Thermo Scientific NanoSpray Flex source using the Rapid Column Housing Unit (1st Generation). The article includes a video instructions on ...
This article demonstrates how to install an Aurora CSI column to your UHPLC system with Bruker CaptiveSpray 2, CaptiveSpray Ultra, and CaptiveSpray Ultra 2 source using a Bruker Column Toaster 2nd Generation. It includes step-by-step guidance from ...