Application Performance Examples
Expected performance of 5cm Aurora Rapid 5x75 column
A Hela tryptic digest (1 ng) was separated on a 5cm Aurora Rapid 5x75 column using a Thermo Scientific Vanquish Neo UHPLC and TSQ Quantis across a 15 min sample gradient at 100 nl/min. The extracted ion chromatograms for 7 selected peptides from the ...
Expected performance of 5cm Aurora Rapid 5x150 column
Tryptic digested Human Plasma was injected and separated using our 180 samples per day method and data dependent acquisition using PASEF resulting in >230 protein identifications (>1,600 unique peptide identifications) [Fig1, Fig 2] with highly ...
Expected performance of 15cm Aurora Elite column
The below graphs represent identified unique proteins, peptides, and average Full Width at Half Maximum (FWHM) from 80ng Hela tryptic digest on a 15cm Aurora Elite column using different gradient lengths. Samples were run on a Waters M-class UPLC ...
Expected performance of 25cm Aurora Ultimate column
The below graphs represent identified unique proteins, peptides, and average Full Width at Half Maximum (FWHM) from 200ng Hela tryptic digest on a 25cm Aurora Ultimate column using different gradient lengths. Samples were run on a Waters M-class UPLC ...
Expected peak widths using the 25cm Aurora Ultimate column
The below plot demonstrates FWHM (seconds) from 200ng Hela cell tryptic digest using different gradients lengths.
Expected peak widths using the 5cm Aurora Rapid 5x150 column
The below boxplot demonstrates FWHM (seconds) from tryptic digested Human Plasma using our 180 samples per day method. The white line indicates median time.
Expected peak widths using the 15cm Aurora Elite column
The below plot demonstrates FWHM (seconds) from 80ng Hela cell tryptic digest using different gradients lengths.
Demonstration of consistent results over multiple experiments and hundreds of injections
80 ng of HeLa peptide was injected and separated on a single 15cm Aurora Elite column across eight proteomics experiments. Sample types include: Cell Lysates Immunoprecipitated Samples Plasma Samples Urine Samples Cell Surface Capture This ...
Application Notes
Visit our Resource Hub to access the latest Application Notes, Technical Notes, and Scientific Posters, to help you get the most out of our products.
Demonstration of consistent results using the 5cm Aurora Rapid150 column
Tryptic digested Hela cell lysate was injected and separated using our 180 samples per day method and data dependent acquisition using PASEF to compare the peptide intensities across samples. This was achieved under the following conditions: Column: ...
Performance of Aurora Series columns with phosphoproteomic samples
Our columns perform very well with phosphoproteomic samples and we have a number of customers which have successfully used our columns for phosphopeptide samples. Below are a few examples of phosphoproteomic publications that have used our columns: ...
Demonstration of consistent peak widths across multiple samples using Aurora Series columns
The below boxplot demonstrates FWHM (seconds) from phospho-peptide (TiO2) enriched samples across 70 injections using a 25cm Aurora Ultimate column. This demonstrates consistent peak widths across multiple samples using Aurora Series columns. Tryptic ...
Expected phosphopeptide identifications using Aurora Series columns
Counts shown on the Y-axis below represent the number of unique phosphorylated peptides identified across 4 replicate runs. These results were achieved under the following conditions: Column: 25cm Aurora Ultimate column Sample: 500ng ...
Performance comparison between Aurora Series columns and competitors
The data in the bar graph below was achieved under the following conditions: Competitor 1 column: 50cm X 75μm Commercial solution without integrated tip Competitor 2 column: 50cm X 75μm Commercial solution with integrated emitter tip IonOpticks ...
BPC: Typical results for Hela Tryptic Digest for 25cm Ultimate, 15cm Elite, and 5cm Rapid Aurora Series columns
The below base peak chromatogram represents typical results of a Hela cell tryptic digest with a 25cm Aurora Series column. This was achieved under the following conditions: Column: 25cm Aurora Ultimate column Sample: 200ng Hela cell tryptic digest ...
Demonstration of stable retention time with Aurora Series columns
A HeLa tryptic digest sample was injected on 6 columns from independent production batches. The retention times for 15 selected peptides were analysed to demonstrate the intra- and inter-column retention time stability for Aurora Series columns. The ...