Expected peak widths using the 5cm Aurora Rapid 5x150 column
Expected peak widths using the 5cm Aurora Rapid 5x150 column
The below boxplot demonstrates FWHM (seconds) from tryptic digested Human Plasma using our 180 samples per day method. The white line indicates median time.
The below boxplot demonstrates FWHM (seconds) from phospho-peptide (TiO2) enriched samples across 70 injections using a 25cm Aurora Ultimate column. This demonstrates consistent peak widths across multiple samples using Aurora Series columns. Tryptic ...
A Hela tryptic digest (1 ng) was separated on a 5cm Aurora Rapid 5x75 column using a Thermo Scientific Vanquish Neo UHPLC and TSQ Quantis across a 15 min sample gradient at 100 nl/min. The extracted ion chromatograms for 7 selected peptides from the ...
Temperature & flow rate We tested flow rates ranging from 50 to 2000 nl/min on the 5 cm Aurora Rapid 5×75 column, all of which produced excellent results. At these flow rates and room temperature, the back pressure from the column should remain ...