Frontier 60x75 column – Recommended conditions
Temperature & flow rate A column oven must be installed at the mass spectrometer source for the 60 cm Aurora Frontier column. For typical operation we recommend a flow rate between 100 and 300 nl/min with the column oven set to 60°C. The maximum ...
Ultimate 25x75 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Ultimate 25x75 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...
Ultimate 25x150 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Ultimate 25x150 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...
Elite 15x150 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Elite 15x150 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at optimal ...
Elite 15x75 column – Recommended conditions
Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the Elite 15x75 column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at optimal ...