Trap Column Volume
NanoShield® 5cm X 100μm = 0.4uL
Loading capacity
Suggested: <200ng
Max: 1ug
For optimal results, it is recommended to load <200 ng onto the NanoShield®, although the maximum capacity is 1 µg. Loading capacity is determined by, and should not exceed the capacity of the analytical column. For optimal results, use the lower value between the recommended trap load and the loading capacity of your analytical column. For more information on loading capacity, see Sample Loading.
Pressure Loading
Depending on your LC system, samples can be loaded using a pressure limit or a set flow rate. The recommended loading pressure for NanoShield® is 1000 bar. Pressures above 1000 bar, if supported by your LC system, may cause peak tailing or other undesirable peak characteristics.
Sample Loading Flow
Loading flow rates depend on your LC system’s pressure limit or can be set between 3-15 µL/min. We recommend loading 5–10 times the trap column volume. For a NanoShield® 5 cm x 100 µm trap, this is ~2.5-5 µL. At 5 µL/min, 10 column volumes takes approximately 1 minute before switching the trap in-line with the analytical column.
Elution, Wash, and Re-equilibration
For elution, use the IonOpticks recommended gradients for your analytical column with the NanoShield® trap. Recommended example gradients for all columns are available here.
Add extra time onto the end of the gradient during the wash and re-equilibration steps to account for any delay introduced by extra volumes due to the NanoShield® trap and capillary lines.