Emitter tip replacements

Emitter tip replacements

The stationary phase in our column is packed directly into the emitter tip to eliminate post-column dead volume. Therefore, the emitter cannot be replaced as a separate item. Provided that samples loaded on to the column are free of salts and other contaminants and voltage is applied to the column during operation, the emitter should perform to a high standard for several months.


    • Related Articles

    • What if there is a fault with my column?

      Our columns are subjected to rigorous quality control procedures under the direct eye of our Senior Scientists. All IonOpticks products are covered by our 100% performance guarantee. Any item not meeting our high-performance expectations due to ...
    • Sample preparation

      Prior to loading samples onto your Aurora Series column, the sample should be de-salted and should not contain any contaminants (salts, detergents, solid particles, etc). Important Loading contaminated samples onto the column may disrupt solvent flow ...
    • Performance comparison between Aurora Series columns and competitors

      The data in the bar graph below was achieved under the following conditions: Competitor 1 column:  50cm X 75μm Commercial solution without integrated tip Competitor 2 column:  50cm X 75μm  Commercial solution with integrated emitter tip IonOpticks ...
    • Removal from a UHPLC

      To maintain optimal performance and extend column lifetime, we recommend keeping the column installed once in use. Avoid removing the column unless performance declines and replacement is required. If removal is necessary, please refer to the Removal ...
    • Low amounts (1mM - 10mM) of HEPES in a sample

      The presence of small concentrations of HEPES should not affect the column chemistry, however, there is a significant risk of fouling the emitter tip leading to spray instability as HEPES is not volatile and may deposit on the tip.