Low amounts (1mM - 10mM) of HEPES in a sample

Low amounts (1mM - 10mM) of HEPES in a sample

The presence of small concentrations of HEPES should not affect the column chemistry, however, there is a significant risk of fouling the emitter tip leading to spray instability as HEPES is not volatile and may deposit on the tip.

    • Related Articles

    • Sample preparation

      Prior to loading samples onto your Aurora Series column, the sample should be de-salted and should not contain any contaminants (salts, detergents, solid particles, etc). Important Loading contaminated samples onto the column may disrupt solvent flow ...
    • Peptides from a sample that are not captured during the loading step

      The stationary phase used in our columns has a lower retentivity of peptides than other stationary phases (peptides are eluted at a lower %ACN than other stationary phases). Therefore, there is a small amount of peptides that will not bind to the ...
    • Standby and idle conditions

      To optimise column lifetime and performance, we recommend that the following conditions are met while your instrument is standby and idle mode: The instrument continues to run at the desired operating pressure The instrument ideally continues to run ...
    • Performance of Aurora Series columns with phosphoproteomic samples

      Our columns perform very well with phosphoproteomic samples and we have a number of customers which have successfully used our columns for phosphopeptide samples. Below are a few examples of phosphoproteomic publications that have used our columns: ...
    • Sample loading

      We recommend the following buffer compositions: Buffer A: 99.9% MilliQ Water, 0.1% formic acid Buffer B: 99.9% Acetonitrile, 0.1% formic acid Please consult with your UHPLC manufacturer to confirm that your UHPLC is compatible with these compositions ...