- Related Articles
- Initial operation and buffer compositions- Once the column is connected to your UHPLC system and is placed inside the source heater or housing, begin operation using 70% buffer B at a flow rate equivalent to the desired gradient flow rate for around 10 minutes or until the pressure is stable ... 
- Standby and idle conditions- To optimise column lifetime and performance, we recommend that the following conditions are met while your instrument is standby and idle mode: The instrument continues to run at the desired operating pressure The instrument ideally continues to run ... 
- Sample loading- We recommend the following buffer compositions: Buffer A: 99.9% MilliQ Water, 0.1% formic acid Buffer B: 99.9% Acetonitrile, 0.1% formic acid Please consult with your UHPLC manufacturer to confirm that your UHPLC is compatible with these compositions ... 
- Removal from a UHPLC- Where possible, we recommend not to remove the column from a UHPLC system before performance falls and the column requires replacement. However, removal may sometimes be required for instrument servicing or maintenance. For more information on ... 
- Emitter tip replacements- The stationary phase in our column is packed directly into the emitter tip to eliminate post-column dead volume. Therefore, the emitter cannot be replaced as a separate item. Provided that samples loaded on to the column are free of salts and other ...