Required accessories for Aurora CSI column operation
Depending on your chosen column length and Instrument setup, the following accessories may be required to ensure optimal performance:
Earth cable
Part No. HVCABLE02
Required for all Aurora CSI column dimensions when used with a Bruker CaptiveSpray source, if the instrument does not already have an earth cable fitted.
Allows the earth connection to be established with the nanoZero® or VHP fitting.
30 mm Grounding Arm for Bruker Column Toaster
Part No. GNDARM01
Required for 5 cm Aurora Rapid columns with a Bruker Column Toaster in a Bruker CaptiveSpray or CaptiveSpray 2 source.
This replaces the plastic grounding arm inside the Bruker Column Toaster and allows the earth connection to be established with the nanoZero®
55 mm Grounding Arm for Bruker Column Toaster
Part No. GNDARM02
Required for 8 cm Aurora Rapid columns with a Bruker Column Toaster in a Bruker CaptiveSpray or CaptiveSpray 2 source.
This replaces the plastic grounding arm inside the Bruker Column Toaster and allows the earth connection to be established with the nanoZero®
Column Heating
Required for heating 60 cm Aurora Frontier CSI columns
- Optional for heating 25 cm Aurora Ultimate CSI, 15 cm Aurora Elite CSI, 8 cm Aurora Rapid 8x75 and 8x150 CSI, and 5 cm Aurora Rapid 5x75 and 5x150 CSI columns.
Note: Operating 25 cm Aurora Ultimate XT or 15 cm Aurora Elite XT columns without a column heater may lead to fluctuating temperatures and increased back pressure at optimal flow rates, generating inconsistent sample separation.
For compatible column heaters with Aurora Series columns, see this support article.
Related Articles
Required accessories for Aurora columns
Depending on your desired column length, one of the accessories listed below is required: High-voltage connection cable Part No. HVCABLE01 Required for 60cm Aurora Frontier, 25cm Aurora Ultimate, and 15cm Aurora Elite columns in a Thermo Nanospray ...
Aurora Series column features
Aurora Series columns are available in four specific compatibility marks, allowing for compatibility with Thermo Scientific, Bruker, and Sciex instruments: Instrument Compatibility Aurora Columns Compatible for use with Thermo Scientific’s Nanospray ...
Required accessories for Aurora XT column operation
Depending on your chosen column length and source setup, the following accessories may be required to ensure optimal performance: HeatSync™ 2-Channel Controller Introducing the HeatSync range of temperature control solutions, which includes: 1 x ...
Required accessories for Aurora XS column operation
Depending on your chosen column length, the following accessories may be required to ensure optimal performance: HeatSync™ 2-Channel Controller Introducing the HeatSync range of temperature control solutions, which includes: 1 x HeatSync™ 2-Channel ...
Aurora CSI column product details
The Aurora CSI range columns are compatible with the Bruker CaptiveSpray source, offering a plug-and-play format tailored to Bruker workflows. Is the Aurora CSI compatible with your workflows? We recommend choosing the Aurora CSI if: You're using the ...