Spray voltage
It is recommended to start with a spray voltage of 1500V and increase over time as required to maintain a stable spray. It is recommended that the spray voltage does not exceed 2500V. Important Exceeding a spray voltage of 2500V may result in damage ...
Example gradients for Aurora Series columns
Sample loading
We recommend the following buffer compositions: Buffer A: 99.9% MilliQ Water, 0.1% formic acid Buffer B: 99.9% Acetonitrile, 0.1% formic acid Please consult with your UHPLC manufacturer to confirm that your UHPLC is compatible with these compositions ...
How to connect the nanoZero® to the nanoViper™ fitting
This video demonstrates how to connect the nanoZero to the Thermo Scientific nanoViper™ fitting: Hold the nanoZero with the spanner. Tighten the nanoViper™ fitting finger tight until you reach a firm stop. A firm stop is the point at which the ...
Aurora Series column features
Aurora Series columns are available in four specific compatibility marks, allowing for compatibility with Thermo Scientific, Bruker, and Sciex instruments: Instrument Compatibility Aurora Columns Compatible for use with Thermo Scientific’s Nanospray ...