Required accessories for Gen3 Aurora CSI columns

Required accessories for Gen3 Aurora CSI columns

Depending on your desired column length and instrument setup, the following accessories may be required:



Earth cable
  1. Part No. HVCABLE02
  2. Required for all Aurora CSI column dimensions in a Bruker CaptiveSpray source if an existing earth cable is not fitted to the instrument
  3. Allows the earth connection to be established with the nanoZero® or VHP fitting




30 mm Grounding Arm for Bruker Column Toaster
  1. Part No. GNDARM01
  2. Required for 5 cm Aurora Rapid columns with a Bruker Column Toaster in a Bruker CaptiveSpray or CaptiveSpray 2 source.
  3. This replaces the plastic grounding arm inside the Bruker Column Toaster and allows the earth connection to be established with the nanoZero® 


55 mm Grounding Arm for Bruker Column Toaster
  1. Part No. GNDARM02
  2. Required for 8 cm Aurora Rapid columns with a Bruker Column Toaster in a Bruker CaptiveSpray or CaptiveSpray 2 source.
  3. This replaces the plastic grounding arm inside the Bruker Column Toaster and allows the earth connection to be established with the nanoZero® 



Column heater
  1. Required for heating 60cm Aurora Frontier CSI columns
  2. Optional for heating 25cm Aurora Ultimate CSI, 15cm Aurora Elite CSI, 5cm Aurora Rapid150 CSI, and 5cm Aurora Rapid75 CSI columns. Please note that operation of 25cm Aurora Ultimate and 15cm Aurora Elite columns without a column heater may result in fluctuating operating temperatures and higher back pressures at optimal flow rates, generating inconsistent sample separation.

If you would like to learn more about which column heaters are compatible with Aurora Series columns, please visit this support article.

    • Related Articles

    • Required accessories for Gen3 Aurora columns

      Depending on your desired column length, one of the accessories listed below is required: High-voltage connection cable Part No. HVCABLE01 Required for 60cm Aurora Frontier, 25cm Aurora Ultimate, and 15cm Aurora Elite columns in a Thermo Nanospray ...
    • Required accessories for Gen3 Aurora XT columns

      Depending on your desired column length and source setup, the following accessories may be required: XT Nanospray Flex Adapter Part No. XTFLXAD01 Required if you wish to operate Aurora XT columns with the Thermo Nanospray Flex source IonOpticks ...
    • Required accessories for Gen3 Aurora XS columns

      All Aurora XS columns require the following accessory: Aurora XS Installation Toolkit Part No. XSITK01 Includes an adapter for Aurora XS columns and probe setting tool to assist with installation Depending on your desired column length, the following ...
    • Aurora Series column features

      Aurora Series columns are available in four product ranges, allowing for compatibility with Thermo Scientific, Bruker, and Sciex instruments: Aurora columns Compatible with Thermo Scientific's Nanospray Flex ion source. Aurora XT columns Compatible ...
    • Gen3 Aurora CSI column product details

      A few things to note about the Aurora CSI product range: These columns are compatible with Bruker's CaptiveSpray source only The columns have a pre-fitted CaptiveSpray capillary holder and directly interface with the CaptiveSpray source Aurora CSI ...