Aurora Series column features
Aurora Series columns are available in four product ranges, allowing for compatibility with Thermo Scientific, Bruker, and Sciex instruments: Aurora columns Compatible with Thermo Scientific's Nanospray Flex ion source. Aurora XT columns Compatible ...
Operation of Aurora Series columns - removal from a UHPLC
If possible, do not remove the column from a UHPLC system before the column performance falls and the columns requires replacement, however, removal for instrument servicing and maintenance is sometimes required. Removal of the nanoViper™ Fingertight ...
Operation of Aurora Series columns - spray voltage
It is recommended to start with a spray voltage of 1500V and increase over time as required to maintain a stable spray. It is recommended that the spray voltage does not exceed 2500V. Exceeding a spray voltage of 2500V may result in damage to the ...
Difference between Gen2 and Gen3 Aurora Series columns
Delivering unrivalled coverage, throughput, sensitivity and reproducibility, Gen3 Aurora Series columns embody all of the strengths of previous generations, augmented by three years of intense research, development and industry collaboration. ...