Operation of Aurora Series columns - initial operation

Operation of Aurora Series columns - initial operation

Once the column is connected to your UHPLC system and is placed inside the source heater or housing, begin operation using 70% buffer B at a flow rate equivalent to the desired gradient flow rate for around 10 minutes or until the pressure is stable for several minutes. Voltage should be applied once the mobile phase reaches the emitter tip. It is recommended that at least one gradient is run without sample injection before assessment of column performance using standards.

We recommend the following buffer compositions:
  1. Buffer A: 99.9% MilliQ Water, 0.1% formic acid
  2. Buffer B: 99.9% Acetonitrile, 0.1% formic acid

Please consult with your UHPLC manufacturer to confirm that your UHPLC is compatible with these compositions.

    • Related Articles

    • Operation of Aurora Series columns - sample loading & buffer compositions

      We recommend the following buffer compositions: Buffer A: 99.9% MilliQ Water, 0.1% formic acid Buffer B: 99.9% Acetonitrile, 0.1% formic acid Please consult with your UHPLC manufacturer to confirm that your UHPLC is compatible with these ...
    • Operation of Aurora Series columns - column equilibration

      Before each run the column should be reequilibrated using a minimum of 4 column volumes of 100% buffer A. Column equilibration should be performed below 1000 bar. Column volumes: Aurora Frontier, 60cm X 75μm: 2.65μl Aurora Ultimate, 25cm X 75μm: ...
    • Operation of Aurora Series columns - standby and idle conditions

      To optimise column lifetime and performance, we recommend that the following conditions are met while your instrument is standby and idle mode: The instrument continues to run at the desired operating pressure The instrument ideally continues to run ...
    • Operation of Aurora Series columns - removal from a UHPLC

      If possible, do not remove the column from a UHPLC system before the column performance falls and the columns requires replacement, however, removal for instrument servicing and maintenance is sometimes required. Removal of the nanoViper™ Fingertight ...
    • Operation of Aurora Series columns - sample preparation

      Prior to loading samples onto your Aurora Series column, the sample should be de-salted and should not contain any contaminants (salts, detergents, solid particles, etc). Loading contaminated samples onto the column may disrupt solvent flow or foul ...