Operation of Aurora Series columns - environmental conditions
The laboratory room temperature must be maintained between 15 and 27 °C (59 and 81 °F).
The relative humidity of the operating environment must be between 20 and 80%, with no condensation.
The operating environment of the equipment should be free of vibrations.
Related Articles
Operation of Aurora Series columns - standby and idle conditions
To optimise column lifetime and performance, we recommend that the following conditions are met while your instrument is standby and idle mode: The instrument continues to run at the desired operating pressure The instrument ideally continues to run ...
Operation of Aurora Series columns - removal from a UHPLC
If possible, do not remove the column from a UHPLC system before the column performance falls and the columns requires replacement, however, removal for instrument servicing and maintenance is sometimes required. Removal of the nanoViper™ Fingertight ...
Operation of Aurora Series columns - column temperature
The recommended operating temperature for Aurora Series columns is 40-60°C. The maximum operating temperature is 60°C. For additional recommendations on operating conditions by column dimension, refer to this section of the Help Centre.
Operation of Aurora Series columns - initial operation
Once the column is connected to your UHPLC system and is placed inside the source heater or housing, begin operation using 70% buffer B at a flow rate equivalent to the desired gradient flow rate for around 10 minutes or until the pressure is stable ...
Operation of Aurora Series columns - sample preparation
Prior to loading samples onto your Aurora Series column, the sample should be de-salted and should not contain any contaminants (salts, detergents, solid particles, etc). Loading contaminated samples onto the column may disrupt solvent flow or foul ...