Column volumes and equilibration

Column volumes and equilibration

Before each run, Aurora Series columns should be equilibrated with a minimum of 4 column volumes of 100% buffer A.

Column volumes:
  1. Aurora Frontier, 60 cm X 75μm: 2.65μl
  2. Aurora Ultimate, 25 cm X 75μm: 1.1μl
  3. Aurora Elite, 15 cm X 75μm: 0.66μl
  4. Aurora 8 cm Rapid 8x150, 8 cm X 150μm: 1.41μl
  5. Aurora 8 cm Rapid 8x75, 8 cm X 75um: 0.35μl
  6. Aurora 5 cm Rapid 5x150, 5 cm X 150μm: 0.88μl
  7. Aurora 5 cm Rapid 5x75, 5 cm X 75um: 0.22μl

    • Related Articles

    • Initial operation and buffer compositions

      Once the column is connected to your UHPLC system and is placed inside the source heater or housing, begin operation using 70% buffer B at a flow rate equivalent to the desired gradient flow rate for around 10 minutes or until the pressure is stable ...
    • Sample loading

      We recommend the following buffer compositions: Buffer A: 99.9% MilliQ Water, 0.1% formic acid Buffer B: 99.9% Acetonitrile, 0.1% formic acid Please consult with your UHPLC manufacturer to confirm that your UHPLC is compatible with these compositions ...
    • Standby and idle conditions

      To optimise column lifetime and performance, we recommend that the following conditions are met while your instrument is standby and idle mode: The instrument continues to run at the desired operating pressure The instrument ideally continues to run ...
    • Removal from a UHPLC

      Where possible, we recommend not to remove the column from a UHPLC system before performance falls and the column requires replacement. However, removal may sometimes be required for instrument servicing or maintenance. For more information on ...
    • Column temperature

      Aurora Series columns are designed to operate optimally at 40–60 °C, with a maximum allowable temperature of 60 °C. For more information on column oven compatibility, click here.