How to revive a column that has been blocked by a 'thick' sample

How to revive a column that has been blocked by a 'thick' sample

A thick sample is often the result of precipitate forming after the addition of loading buffer to a sample vial. There are two things you can try to revive the column:
  1. Run 5 - 10+ blank gradients
  2. Run 85% acetonitrile at 400nl/min for an extended period of time
We have tried testing a variety of other solvents to clean up columns but have not discovered anything suitable. If the back pressure remains below the operating maximum of your UHPLC then we normally find that the chromatography isn't significantly affected in these situations so you may be able to continue with additional samples using the same column.
    • Related Articles

    • Sample preparation

      Prior to loading samples onto your Aurora Series column, the sample should be de-salted and should not contain any contaminants (salts, detergents, solid particles, etc). Important Loading contaminated samples onto the column may disrupt solvent flow ...
    • Performance of Aurora Series columns with phosphoproteomic samples

      Our columns perform very well with phosphoproteomic samples and we have a number of customers which have successfully used our columns for phosphopeptide samples. Below are a few examples of phosphoproteomic publications that have used our columns: ...
    • Standby and idle conditions

      To optimise column lifetime and performance, we recommend that the following conditions are met while your instrument is standby and idle mode: The instrument continues to run at the desired operating pressure The instrument ideally continues to run ...
    • Expected phosphopeptide identifications using Aurora Series columns

      Counts shown on the Y-axis below represent the number of unique phosphorylated peptides identified across 4 replicate runs.  These results were achieved under the following conditions: Column:  25cm Aurora Ultimate column Sample:  500ng ...
    • BPC: Typical results for Hela Tryptic Digest for 25cm Ultimate, 15cm Elite, and 5cm Rapid Aurora Series columns

      The below base peak chromatogram represents typical results of a Hela cell tryptic digest with a 25cm Aurora Series column. This was achieved under the following conditions: Column: 25cm Aurora Ultimate column Sample: 200ng Hela cell tryptic digest ...