Expected back pressure, 25cm Aurora Ultimate column

Expected back pressure, 25cm Aurora Ultimate column

The below graph depicts the expected backpressure of a 25cm Aurora Ultimate column, for a given flow rate at both room temperature and 40°C.

The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid.

The Aurora Series column and nanoZero® fitting can both handle pressures >1700 bar (Generation 3) and >1200bar (Generation 2).



    • Related Articles

    • Expected back pressure, 15cm Aurora Elite column

      The below graph depicts the expected backpressure of a 15cm Aurora Elite column, for a given flow rate at both room temperature and 40°C. The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid. The Aurora Series column and ...
    • Expected back pressure, 5cm Aurora Rapid75 column

      The below graph depicts the expected backpressure of a 5cm Aurora Rapid75 column, for a given flow rate at both room temperature and 40°C. The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid. The Aurora Series column and ...
    • Expected back pressure, 60cm Aurora Frontier column

      The below graph depicts the expected backpressure of a 60cm Aurora Frontier column, for a given flow rate at 60°C. This graph was recorded after 50 sample injections. The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid. The ...
    • Expected back pressure, 5cm Aurora Rapid150 column

      The below graph depicts the expected backpressure of a 5cm Aurora Rapid150 column, for a given flow rate at both room temperature and 40°C. The buffer composition is 79.9% water, 20% acetonitrile, 0.1% formic acid. The Aurora Series column and ...
    • Recommended conditions for 25cm Aurora Ultimate column

      Temperature & flow rate We recommend the use of a column oven installed at the mass spectrometer source for the 25cm Aurora Ultimate column. Operation without a column oven may result in fluctuating operating temperatures and higher back pressures at ...