Popular Articles
Spray voltage
It is recommended to start with a spray voltage of 1500V and increase over time as required to maintain a stable spray. It is recommended that the spray voltage does not exceed 2500V. Important Exceeding a spray voltage of 2500V may result in damage ...
Example gradients for Aurora Series columns
How to connect the nanoZero® to the nanoViper™ fitting
This video demonstrates how to connect the nanoZero to the Thermo Scientific nanoViper™ fitting: Hold the nanoZero with the spanner. Tighten the nanoViper™ fitting finger tight until you reach a firm stop. A firm stop is the point at which the ...
Aurora Series column features
Aurora Series columns are available in four specific compatibility marks, allowing for compatibility with Thermo Scientific, Bruker, and Sciex instruments: Instrument Compatibility Aurora Columns Compatible for use with Thermo Scientific’s Nanospray ...
Removal from a UHPLC
To maintain optimal performance and extend column lifetime, we recommend keeping the column installed once in use. Avoid removing the column unless performance declines and replacement is required. If removal is necessary, please refer to the Removal ...